MR imaging-guided percutaneous nephrostomy and use of MR-compatible catheters in the nondilated porcine urinary tract.

Nolte-Ernsting C C, Bücker A, Neuerburg J M, Glowinski A, Adam G B, Günther R W

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To investigate technique and practicability of MR-guided percutaneous nephrostomy (MRPCN) and to test magnetic resonance (MR)-compatible catheters inside the urinary tract.

MATERIALS AND

METHODS

In 10 healthy pigs, a percutaneous nephrostomy tube was placed into the nonobstructed pelvicaliceal system with use of exclusive MR guidance with a standard 1.5-T magnet. The urinary tract was visualized by intravenous injection of Gd-DTPA in combination with low-dose furosemide. The procedure was controlled with use of a T1-weighted turbo gradient-echo sequence in two orthogonal planes. The equipment for MRPCN included an 18-gauge MR-compatible puncture needle, a nitinol guide wire, and different 5-F MR catheters.

RESULTS

In all 10 animals, the puncture needle was safely directed into the nondilated target calix. Slight deviations of the needle were detected on both MR image planes, which enabled immediate correction. This technique achieved a "first attempt" puncture of the targeted calix in each animal. MR images accurately demonstrated the dysprosium labelled tip of the different inserted catheters. It proved essential to inject a gadolinium-insaline solution via these catheters to preserve the endoluminal contrast enhancement as long as necessary. Balloon catheters were directed and inflated inside the ureter under exclusive MR guidance. Complications such as perforation and leakage were visualized by MR imaging.

CONCLUSIONS

MRPCN is a promising technique for puncturing the pelvicaliceal system. The ability to successfully enter the urinary tract, even when it is nondilated, underscores the accuracy achievable with multiplanar MR imaging.

Для цитирования: Nolte-Ernsting C C, Bücker A, Neuerburg J M, Glowinski A, Adam G B, Günther R W. MR imaging-guided percutaneous nephrostomy and use of MR-compatible catheters in the nondilated porcine urinary tract.. Journal of vascular and interventional radiology : JVIR. 1999. DOI: 10.1016/s1051-0443(99)70236-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.