Precision and accuracy of cross-sectional area measurements used to measure coronary endothelial function with spiral MRI.
СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Schär Michael, Soleimanifard Sahar, Bonanno Gabriele, Yerly Jérôme, Hays Allison G, Weiss Robert G
Аннотация
PURPOSE: Coronary endothelial function (CEF) reflects vascular health and conventional invasive CEF measures predict cardiovascular events. MRI can now noninvasively measure CEF by quantifying coronary artery cross-sectional area changes in response to isometric handgrip exercise, an endothelial-dependent stressor. Area changes (10 to 20% in healthy; 2 to -12% in impaired vessels) are only a few imaging voxels because of MRI's limited spatial resolution. Here, with numerical simulations and phantom studies, we test whether Fourier interpolation enables sub-pixel area measurement precision and determine the smallest detectable area change using spiral MRI.
METHODS: In vivo coronary SNR with the currently used CEF protocol at 3T was measured in 7 subjects for subsequent in vitro work. Area measurements of circular vessels were simulated by varying partial volume, vessel diameter, voxel size, SNR, and Fourier interpolation factor. A phantom with precision-drilled holes (diameters 3-3.42 mm) was imaged 10 times with the current CEF protocol (voxel size, Δx = 0.89 mm) and a high-resolution protocol (Δx = 0.6 mm) to determine precision, accuracy, and the smallest detectable area changes.
RESULTS: In vivo coronary SNR ranged from 30-76. Eight-fold Fourier interpolation improved area measurement precision by a factor 6.5 and 4.9 in the simulations and phantom scans, respectively. The current CEF protocol can detect mean area changes of 4-5% for SNR above 30, and 3-3.5% for SNR above 40 with a higher-resolution protocol.
CONCLUSION: Current CEF spiral MRI with in vivo SNR allows detection of a 4-5% area change and Fourier interpolation improves precision several-fold to sub-voxel dimensions.
METHODS: In vivo coronary SNR with the currently used CEF protocol at 3T was measured in 7 subjects for subsequent in vitro work. Area measurements of circular vessels were simulated by varying partial volume, vessel diameter, voxel size, SNR, and Fourier interpolation factor. A phantom with precision-drilled holes (diameters 3-3.42 mm) was imaged 10 times with the current CEF protocol (voxel size, Δx = 0.89 mm) and a high-resolution protocol (Δx = 0.6 mm) to determine precision, accuracy, and the smallest detectable area changes.
RESULTS: In vivo coronary SNR ranged from 30-76. Eight-fold Fourier interpolation improved area measurement precision by a factor 6.5 and 4.9 in the simulations and phantom scans, respectively. The current CEF protocol can detect mean area changes of 4-5% for SNR above 30, and 3-3.5% for SNR above 40 with a higher-resolution protocol.
CONCLUSION: Current CEF spiral MRI with in vivo SNR allows detection of a 4-5% area change and Fourier interpolation improves precision several-fold to sub-voxel dimensions.