Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging.

Humphreys Mandy, Guerrero Urbano M Teressa, Mubata Cefas, Miles Elizabeth, Harrington Kevin J, Bidmead Margaret, Nutting Christopher M

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate set-up reproducibility of a cabulite shell and determine CTV-PTV margins for head and neck intensity-modulated-radiotherapy.

MATERIALS AND

METHODS

Twenty patients were entered into the study. A total of 354 anterior and lateral isocentric electronic portal images (EPIs) were compared to simulator reference images.

RESULTS

About 94% of all translational displacements were < or =3 mm, and 99% < or =5 mm. The overall systematic error was 0.9 mm (+/-1.0SD) in the Right-Left, 0.7 mm (+/-0.9SD) in the Superior-Inferior and -0.02 mm (+/-1.1SD) in the Anterior-Posterior directions. The corresponding SDs of the random errors were +/-0.4, +/-0.6 and +/-0.7 mm. The estimated margins required from CTV-PTV were calculated according to the Van Herk formula was 2.9, 2.6 and 3.3 mm, respectively.

CONCLUSIONS

This head and neck immobilisation system is of sufficient accuracy for its use with IMRT treatments and a 3 mm CTV-PTV margin has been adopted.

Для цитирования: Humphreys Mandy, Guerrero Urbano M Teressa, Mubata Cefas, Miles Elizabeth, Harrington Kevin J, Bidmead Margaret, Nutting Christopher M. Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2005. DOI: 10.1016/j.radonc.2005.06.039.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.