Identification of Rotors during Human Atrial Fibrillation Using Contact Mapping and Phase Singularity Detection: Technical Considerations.

Kuklik Pawel, Zeemering Stef, van Hunnik Arne, Maesen Bart, Pison Laurent, Lau Dennis H, Maessen Jos, Podziemski Piotr, Meyer Christian, Schaffer Benjamin, Crijns Harry, Willems Stephan, Schotten Ulrich

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To explore technical challenges of phase singularity (PS) mapping during atrial fibrillation (AF) using direct contact electrograms.

METHODS

AF mapping was performed in high-density epicardial recordings of human paroxysmal (PAF) or persistent (PersAF) (N = 20 pts) AF with an array of 16 × 16 electrodes placed on atrial epicardium. PS points were detected using subsets of electrodes forming rings of varying sizes.

RESULTS

PS detection using a 2 × 2 electrode ring identified 0.88 ± 1.00 PS/s in PAF group and 3.91 ± 2.51 per s in PersAF group (p < 0.001) in 2.4 × 2.4 cm mapping area. All detected PS had a short lifespan with the longest being 1100 ms (6.8 rotations). Exploration of the PS detection in a numerical model demonstrated that at least eight electrodes are required to avoid frequent false positive PS detection due to chance. Application of a detection grid consisting a double ring of electrodes (2 × 2 and 4 × 4 rings) decreased the number of false positive detections. The double ring was more resilient to electrode swapping (with just three instances of false positives versus 4380 false positives using 2 × 2 ring).

CONCLUSIONS

The number of detected rotors critically depends upon the parameters of the detection algorithm, especially the number of electrodes used to detect PS. Based on our results, we recommend double ring comprised of 2 × 2 and 4 × 4 grid of electrodes for robust rotor detection.

SIGNIFICANCE: Great methodological care has to be taken before equating detected PS with rotating waves and using PS detection algorithms to guide catheter ablation of AF.

Для цитирования: Kuklik Pawel, Zeemering Stef, van Hunnik Arne, Maesen Bart, Pison Laurent, Lau Dennis H, Maessen Jos, Podziemski Piotr, Meyer Christian, Schaffer Benjamin, Crijns Harry, Willems Stephan, Schotten Ulrich. Identification of Rotors during Human Atrial Fibrillation Using Contact Mapping and Phase Singularity Detection: Technical Considerations.. IEEE transactions on bio-medical engineering. 2016. DOI: 10.1109/tbme.2016.2554660.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.