Scandinavian journal of clinical and laboratory investigation · 2009
Determination of salivary cortisol by liquid chromatography-tandem mass spectrometry.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Turpeinen Ursula, Välimäki Matti Juhani, Hämäläinen Esa
Аннотация
BACKGROUND: Late evening salivary cortisol concentrations are increasingly used as a screening test in suspected Cushing's syndrome partly because of easy sample collection. The cortisol immunoassays are prone to interference by cross-reacting steroids and therefore there is a need for improvement. The high specificity of an LC-MS assay provides a solution to the problem.
METHODS: Our liquid chromatography-tandem mass spectrometric (LC-MS/MS) analysis utilizes only 0.1 ml of saliva. The samples were extracted with dichloromethane. The extract was evaporated to dryness and cortisol was analysed by LC-MS/MS operating in the negative mode ESI after separation on a reversed-phase column.
RESULTS: The calibration curves for analysis of salivary cortisol exhibited consistent linearity and reproducibility in the range of 0.5-20 nmol/L. Interassay CVs were 4.3-11% at cortisol concentrations of 0.6-14 nmol/L. The lower limit of quantitation (LOQ) was 70 pmol/L (signal to noise ratio=10). The mean recovery of the analyte added to saliva samples ranged from 95-106%. The upper limit of the reference range (95%) was 3.0 nmol/L.
CONCLUSIONS: Our method is rapid, sensitive and simple to perform with a routine LC-MS/MS spectrometer.
METHODS: Our liquid chromatography-tandem mass spectrometric (LC-MS/MS) analysis utilizes only 0.1 ml of saliva. The samples were extracted with dichloromethane. The extract was evaporated to dryness and cortisol was analysed by LC-MS/MS operating in the negative mode ESI after separation on a reversed-phase column.
RESULTS: The calibration curves for analysis of salivary cortisol exhibited consistent linearity and reproducibility in the range of 0.5-20 nmol/L. Interassay CVs were 4.3-11% at cortisol concentrations of 0.6-14 nmol/L. The lower limit of quantitation (LOQ) was 70 pmol/L (signal to noise ratio=10). The mean recovery of the analyte added to saliva samples ranged from 95-106%. The upper limit of the reference range (95%) was 3.0 nmol/L.
CONCLUSIONS: Our method is rapid, sensitive and simple to perform with a routine LC-MS/MS spectrometer.