Organ and effective dose evaluation in coronary angiography by using a 320 MDCT based on in-phantom dose measurements with TLDs.

Lu Heqing, Zhuo Weihai, Xu Bing, Wang Minjie

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

The purpose of this study is to estimate the organ and effective dose (E) from computed tomography coronary angiography (CTCA) on a 320-MDCT scanner. Radiation dose was estimated for the prospectively ECG-gated CTCA in a male phantom. A total of 451 chips of thermoluminescent dosimeter were implanted in the phantom for measuring the organ doses. The effective doses were calculated using measured organ doses and the tissue-weighting factors. The dose length product (DLP) values were recorded and used to develop the conversion coefficient k = 0.017 mSv•(mGy•cm)(-1) (E/DLP). In a 3-beat acquisition, the organ doses ranged from 0.24 to 71.55 mGy, and the doses in breast, bone surface, oesophagus, and lung were higher than 20 mGy. The effective doses in 2-beat and 3-beat acquisition were estimated to be 14.3 and 24.3 mSv. More beats of acquisition led to higher radiation dose. The reported k values for chest CT scan can be used to roughly estimate the E value from CTCA for 320 MDCT.

Для цитирования: Lu Heqing, Zhuo Weihai, Xu Bing, Wang Minjie. Organ and effective dose evaluation in coronary angiography by using a 320 MDCT based on in-phantom dose measurements with TLDs.. Journal of radiological protection : official journal of the Society for Radiological Protection. 2015. DOI: 10.1088/0952-4746/35/3/597.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.