Increasing potential risks of contamination from repetitive use of endoscope.

Lee Dae-Hyung, Kim Dong Bin, Kim Hyun Yong, Baek Hyun Sook, Kwon Soon-Young, Lee Mi Hee, Park Jong-Chul

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Medical devices inserted into the human body can be divided into single-use devices and reusable medical devices (RMDs) depending on the method of use. RMDs are applied to >1 patient, and users reprocess them prior to repeat use. Therefore, reprocessing of RMDs has become more difficult, and issues related to infection risk have become more prevalent.

METHODS

Research was performed to compare differences in surface alterations between not-aged (NA) samples and accelerated-aging (AA) samples. NA and AA samples were exposed to potential contaminants and treated with the same cleaning conditions. Then the residual contaminants were analyzed, and adhesion characteristics were investigated.

RESULTS

Significant differences between the 2 sample groups were observed in the pattern of physical surface alterations. Similar to the endoscope for clinical use, a deep crack was found in the surface of the AA sample. According to the binding affinity test, highly likely potential contaminants were found more frequently on AA samples.

CONCLUSION

The results suggest the necessity of limiting the duration of time that RMDs may be used, considering the potential risk of infection from repetitive use of RMDs, and also raised the possibility of applying this simulation model of AA. This methodology can be applied to various RMDs, including endoscopes.

Для цитирования: Lee Dae-Hyung, Kim Dong Bin, Kim Hyun Yong, Baek Hyun Sook, Kwon Soon-Young, Lee Mi Hee, Park Jong-Chul. Increasing potential risks of contamination from repetitive use of endoscope.. American journal of infection control. 2015. DOI: 10.1016/j.ajic.2015.01.017.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.