MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/0952-8180(91)90005-8

Медицинская библиотека · Journal of clinical anesthesia · 1991

Duration of carbon dioxide absorption by soda lime at low rates of fresh gas flow.

Ohrn M, Gravenstein N, Good M L

СтатьяПеревод в очереди

Аннотация

STUDY

OBJECTIVE

To determine the impact of a low fresh gas flow rate on the duration of carbon dioxide (CO2) absorption by soda lime.

DESIGN

Nonclinical, experimental.

SETTING

Experimental laboratory.

METHODS

In vitro test with Sodasorb and a semiclosed breathing circle ventilating a test lung with a CO2 inflow of 250 ml per minute. Fresh gas flow rates of 0.25, 0.5, 1, 2, and 4 L/min were studied.

MEASUREMENTS AND MAIN

RESULTS

CO2 was measured at the breathing circuit test lung interface with a mainstream capnometer. Duration of CO2 absorption was determined as the time for the inspired CO2 tension (PICO2) to increase from 0 mm to 7 mm of mercury. The times of this interval were recorded four times for each fresh gas flow rate and compared by analysis of variance; p less than 0.05 was considered significant. Time to soda lime failure was significantly longer at 2 L/min than at 1 L/min fresh gas flow and at 1 L/min than at 0.25 L/min fresh gas flow.

CONCLUSION

Because soda lime color indicators are unreliable, when a semiclosed breathing circle is used at a low rate of fresh gas flow without CO2 monitoring, the CO2 absorbent must be replaced more frequently.

Для цитирования: Ohrn M, Gravenstein N, Good M L. Duration of carbon dioxide absorption by soda lime at low rates of fresh gas flow.. Journal of clinical anesthesia. 1991. DOI: 10.1016/0952-8180(91)90005-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.