Медицинская библиотека · Aerospace medicine and human performance · 2022

A Computer-Based Farnsworth-Munsell 100-Hue (CFM-100) Test in Pilots' Medical Assessments.

Zhang Yinjuan, Ma Jin, Cheng Shan, Hu Wendong

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

This study investigated the effectiveness and identified the cutoff values of the computer-based Farnsworth-Munsell 100-Hue (CFM-100) test for screening color vision deficiencies in the pre-employment examination of civil aviators in China.

METHODS

Firstly, subjects were stratified into normal, color weakness, and color blindness with the Ishihara pseudoisochromatic plate test (IPPT) by two ophthalmologists. Then they randomly completed CFM-100 and Farnsworth-Munsell 100-Hue (FM-100) tests. Total error scores (TES) and the time taken for the CFM-100 and FM-100 were analyzed and the cutoff values for the CFM-100 were determined.

RESULTS

Of 218 subjects, 159 were normal while 59 were diagnosed with dyschromatopsia. The TES of the CFM-100 were congruent with those of the FM-100 (20.0 ± 18.8 vs. 20.6 ± 17.7, 160.9 ± 66.0 vs. 151.1 ± 66.4). The testing time for the CFM-100, however, was less than the FM-100 (10.3 ± 2.8 min vs. 12.9 ± 2.9 min, 7.8 ± 2.5 min vs. 12.6 ± 3.3 min). The correlation coefficient R was 0.93 and Cohen's kappa was 0.89 for the two methods. Further analyses defined 34 as the cutoff value to differentiate excellent from fair color discrimination (sensitivity 58.0%, specificity 94.7%) and 101 as the cutoff value to judge fair vs. poor (sensitivity and specificity both 98.8%) for the CFM-100. The cut-off value was 72 for distinguishing normal from defective color vision (sensitivity 96.6%, specificity 98.7%) and 110 was for distinguishing color weakness from color blindness (sensitivity 97.6%, specificity 97.7%) for the CFM-100.

CONCLUSIONS

The CFM-100 is an effective method for the diagnosis of dyschromatopsia with high sensitivity in screening airline pilots.Zhang Y, Ma J, Cheng S, Hu W. A computer-based Farnsworth-Munsell 100-Hue (CFM-100) test in pilots' medical assessments. Aerosp Med Hum Perform. 2022; 93(4):362-367.

Для цитирования: Zhang Yinjuan, Ma Jin, Cheng Shan, Hu Wendong. A Computer-Based Farnsworth-Munsell 100-Hue (CFM-100) Test in Pilots' Medical Assessments.. Aerospace medicine and human performance. 2022. DOI: 10.3357/amhp.5943.2022.