Evaluation of external ventricular drain performance under hyperbaric conditions: a bench study.

СтатьяОценочное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ramchandani Rahulkumar, Lin Zhiliang C, Devaney Bridget, Jeffree Rosalind L, Tsouras Theo

Аннотация

INTRODUCTION: External ventricular drains (EVDs) are widely used for intracranial pressure (ICP) monitoring and cerebrospinal fluid (CSF) diversion in neurocritical care. Hyperbaric oxygen treatment (HBOT) exposes medical devices to increased ambient pressure; however, limited data exists regarding EVD performance and reliability under hyperbaric conditions.

METHODS: A bench study was conducted to evaluate EVD performance under hyperbaric conditions. A rigid, non-compressible saline-filled column was used to assess pressure measurement stability independent of compliance effects. Additional compliance-based syringe models were explored to assess the feasibility of evaluating unclamped drainage behaviour. A commercially available EVD system (CODMAN EDS 3; Codman and Shurtleff, a Johnson and Johnson company, Raynham, MA, USA) was exposed to stepwise ambient pressures from 101 to 283 kPa (1.0-2.8 atmospheres absolute) in a multiplace hyperbaric chamber. Pressure measurements, drainage behaviour, fluid column stability, and system integrity were assessed.

RESULTS: The CODMAN EDS 3 EVD system maintained structural integrity across tested pressure levels. While pressure measurements were stable in the rigid open glass syringe model, the closed compliance-based models demonstrated pressure variability during chamber pressurisation, including when closed to fluid drainage. Unclamped testing using these models demonstrated inconsistent fluid dynamics, precluding reliable assessment of fluid drainage behaviour under hyperbaric conditions.

CONCLUSIONS: The CODMAN EDS 3 EVD system remained structurally intact during hyperbaric exposure; however, pressure measurements varied in closed models, including when closed to fluid drainage. ICP measurements obtained during HBOT should be interpreted with caution due to potential pressure-related artefacts. Reliable continuous assessment of fluid pressure drainage behaviour remains limited by current bench model constraints. Further research incorporating advanced brain compliance simulation models and carefully designed in-vivo observational studies is required to better characterise EVD behaviour under hyperbaric conditions and to ensure safe clinical use and accurate interpretation of ICP measurements in the hyperbaric environment.