How to reduce the risk of organ injuries during surgical instrument insertion in laparoscopic surgery: Pushing/pressing force analysis using forceps with sensors.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Makiyama Kazuhide, Osaka Kimito, Araki Azumi, Ohtake Shinji, Tatenuma Tomoyuki, Nagasaka Manabu, Yamada Takahiro, Yao Masahiro
Аннотация
INTRODUCTION: In laparoscopic surgery, surgical instruments are inserted from a trocar to the target organ in a blind fashion, which carries a risk of organ injury. To clarify the risks associated with surgical instrument insertion, we measured grip strength and pushing/pressing force during surgical instrument insertion in laparoscopic surgery.
METHODS: Using forceps with sensors inside a trocar, 10 urologists performed a laparoscopic procedure in pigs, in which they were asked to touch the abdominal wall. The surgeons closed their eyes during the procedure and stopped moving the forceps when they felt them come into contact with the abdominal wall. They were ordered to grip the forceps strongly or softly and to move them rapidly or slowly during the procedure. Grip strength and the pushing/pressing force when the forceps hit the abdominal wall were measured and analyzed.
RESULTS: The mean pushing/pressing force when the surgeons gripped the forceps strongly and moved them rapidly (strong/rapid), strongly/slowly, softly/rapidly, and softly/slowly were 2.8, 2.0, 1.7, and 1.1 N, respectively. The pushing/pressing force was significantly greater when the surgeons gripped the forceps strongly, regardless of the forceps speed (P < .001). The pushing/pressing force was significantly greater when the surgeons moved the forceps rapidly, regardless of grip strength (P < .001).
CONCLUSIONS: When surgeons insert laparoscopic instruments through trocars, the instruments should be gripped softly and moved slowly to avoid organ injuries.
METHODS: Using forceps with sensors inside a trocar, 10 urologists performed a laparoscopic procedure in pigs, in which they were asked to touch the abdominal wall. The surgeons closed their eyes during the procedure and stopped moving the forceps when they felt them come into contact with the abdominal wall. They were ordered to grip the forceps strongly or softly and to move them rapidly or slowly during the procedure. Grip strength and the pushing/pressing force when the forceps hit the abdominal wall were measured and analyzed.
RESULTS: The mean pushing/pressing force when the surgeons gripped the forceps strongly and moved them rapidly (strong/rapid), strongly/slowly, softly/rapidly, and softly/slowly were 2.8, 2.0, 1.7, and 1.1 N, respectively. The pushing/pressing force was significantly greater when the surgeons gripped the forceps strongly, regardless of the forceps speed (P < .001). The pushing/pressing force was significantly greater when the surgeons moved the forceps rapidly, regardless of grip strength (P < .001).
CONCLUSIONS: When surgeons insert laparoscopic instruments through trocars, the instruments should be gripped softly and moved slowly to avoid organ injuries.