DOI 10.1016/j.ihj.2015.08.024

Медицинская библиотека · Indian heart journal · 2015

Perclose Closure Device breakage and embolization during deployment followed by retrieval with snare.

Kang Kevin, Ferraro Robert, Petrella Richard, Kang Gurjaipal

Клинический случайСтатьяПеревод в очереди

Аннотация

Perclose Closure Device (Abbott-Vascular) was attempted for femoral artery (FA) access site closure by an experienced operator. The device was felt to snap as it was pushed through the previously scarred skin and subcutaneous tissue. Perclose had broken into two pieces with the proximal portion outside the body and the distal portion embolizing to lie in the FA to distal aorta. The left FA was accessed through 8F sheath and 4-8mm Basket snare passed through a 6F multipurpose guiding catheter. The tip of the broken Perclose was caught with the snare and pulled out. The Perclose has a distal flexible sheath that is overmolded directly onto the proximal rigid guide. This junction of sheath and guide may be a weak point. Accidental pressure applied at this weak point in any Perclose device during deployment can cause breakage. Awareness of this rare complication is important for safety.

Для цитирования: Kang Kevin, Ferraro Robert, Petrella Richard, Kang Gurjaipal. Perclose Closure Device breakage and embolization during deployment followed by retrieval with snare.. Indian heart journal. 2015. DOI: 10.1016/j.ihj.2015.08.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.