Active push deployment technique improves stent/vessel-wall interaction in endovascular treatment of acute stroke with stent retrievers.

Wiesmann Martin, Brockmann Marc-Alexander, Heringer Sarah, Müller Marguerite, Reich Arno, Nikoubashman Omid

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The optimal interaction between stent struts and thrombus is crucial for successful revascularization in endovascular stroke therapy with stent retrievers. Deploying the stent retriever by actively pushing it into the thrombus increases the radial force with which the stent struts expand into the thrombus.

OBJECTIVE

To examine the active push deployment (APD) technique in an in vitro model and present our clinical experience with this technique.

METHODS

In an in vitro experiment we investigated the configuration of a Solitaire and a Trevo ProVue device (both 4×20 mm), depending on whether the devices were deployed using the APD technique or simple unsheathing. We retrospectively assessed the effectiveness and safety of this technique by analyzing 130 patients with large vessel occlusions (carotid T or M1 segment of the middle cerebral artery), who received endovascular treatment with a Trevo device (4×20 mm) that was deployed using the APD technique.

RESULTS

In vitro experiment: the APD technique improved apposition of the devices to the vessel wall. There was widening of 30% (Trevo) and 19% (Solitaire) at the cost of a shortening of 5% and 4%, respectively, when the devices were deployed in a carotid T model. Clinical study: the revascularization rate (Thrombolysis in Cerebral Infarction ≥2b) with the Trevo device was 90%. There were no retriever-associated dissections or perforations in 278 retrieval maneuvers.

CONCLUSIONS

The APD technique improves apposition of the tested devices to the vessel wall. The widening effect comes at the cost of minimal shortening of the devices. Our clinical experience shows that using the APD technique to deploy the Trevo device is effective and safe.

Для цитирования: Wiesmann Martin, Brockmann Marc-Alexander, Heringer Sarah, Müller Marguerite, Reich Arno, Nikoubashman Omid. Active push deployment technique improves stent/vessel-wall interaction in endovascular treatment of acute stroke with stent retrievers.. Journal of neurointerventional surgery. 2016. DOI: 10.1136/neurintsurg-2016-012322.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.