[Combined use of bare stent and coils in the treatment of aortic dissection with distal tear at celiac trunk].
Аннотация
OBJECTIVE
To evaluate a novel method of combining use of bare stent and coils in the treatment of aortic dissection with distal tear at celiac trunk.
METHODS
From April 2007 to August 2010, two cases of Stanford type B dissections underwent initially endovascular stent-grafting to seal proximal tears while distal tears at celiac trunks failed to seal spontaneously. Bare stents were then implanted into celiac trunks crossing tears. And an appropriate amount of coils were pushed into false lumen via a catheter passing through the holes of stents for maintaining the patency of celiac trunk and sealing tears.
RESULTS
Both procedures succeeded. False lumen was indistinct on angiography while celiac trunks were patent. No complication occurred. After the follow-up periods of 5 and 2 years respectively, false lumen disappeared and celiac arteries became patent.
CONCLUSION
The novel method of combining use of bare stent and coils is technically simple. It required a simple puncture. Major traumas of open and hybrid operations are avoided. It is also indicated for cases of tears at renal or superior mesenteric arteries. As a supplemental method for conventional TEVAR, radical cure may be achieved for some complicated Stanford B dissections.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.