DOI 10.1097/01.mlg.0000187405.57567.ae

Медицинская библиотека · The Laryngoscope · 2006

Improving vestibular evoked myogenic potential reliability by using a blood pressure manometer.

Vanspauwen Robby, Wuyts Floris L, Van de Heyning Paul H

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE/HYPOTHESIS: To improve the reliability of vestibular evoked myogenic potentials (VEMP), we propose a feedback method making use of a readily available blood pressure manometer with inflatable cuff to control the sternocleidomastoid muscle (SCM) contraction.

STUDY

DESIGN

Prospective study.

MATERIALS AND

METHODS

The feedback method for SCM contraction consisted of subjects pushing with their jaw against the hand-held inflated cuff to generate a specified cuff pressure. This pressure level was monitored by subject and investigator. First, we tested in a group of healthy subjects whether there was a learning or fatigue effect during four successive SCM contraction runs when making use of the feedback method. Then, we investigated the mean rectified voltage (MRV) values of the SCM. Next, we examined the hypothesis that the MRV values of the SCM contraction measured before and simultaneously with the VEMP test were equal when using this feedback method. Finally, we compared the VEMP amplitude variability in two circumstances: with and without the feedback method.

RESULTS

There was no learning or fatigue effect, and the MRV values measured before the VEMP were not significantly different from those measured during the VEMP test. The VEMP amplitude variability was significantly lower when applying the feedback method than when no feedback method was used.

CONCLUSION

This feedback method improves significantly the VEMP amplitude reliability.

Для цитирования: Vanspauwen Robby, Wuyts Floris L, Van de Heyning Paul H. Improving vestibular evoked myogenic potential reliability by using a blood pressure manometer.. The Laryngoscope. 2006. DOI: 10.1097/01.mlg.0000187405.57567.ae.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.