Gastrointestinal endoscopy · 2000
The measurement of forces exerted during colonoscopy.
Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Appleyard M N, Mosse C A, Mills T N, Bell G D, Castillo F D, Swain C P
Аннотация
BACKGROUND: The widely varying forces used at colonoscopy have not been measured. An electronic device was designed to measure the forces exerted by the clinician on the endoscope during colonoscopy.
METHODS: The device featured a handle designed in the shape of a hinged split cylinder that could be locked around the endoscope but readily moved up and down the insertion tube as the colonoscopy proceeded. This cylinder contained strain-gauges arranged so that the forces transmitted could be accurately measured. The device recorded the torque forces in addition to the push and pull forces exerted during diagnostic colonoscopy.
RESULTS: In a series of 21 colonoscopies in 20 patients: peak pushing force = 4.4 kg, pulling force = -1.8 kg, anti-clockwise torque = 1.0 Newton meters, clockwise torque = 0.8 Newton meters. Percentage time force greater than 1 kg = 5%. Peak anal insertion force = 1.8 kg.
CONCLUSIONS: These measurements represent the first accurate measurements of the forces exerted during colonoscopy. Reducing the force during colonoscopy is likely to diminish pain and reduce the risk of perforation. A knowledge of these forces may also help with the design of new instruments and models for teaching or research.
METHODS: The device featured a handle designed in the shape of a hinged split cylinder that could be locked around the endoscope but readily moved up and down the insertion tube as the colonoscopy proceeded. This cylinder contained strain-gauges arranged so that the forces transmitted could be accurately measured. The device recorded the torque forces in addition to the push and pull forces exerted during diagnostic colonoscopy.
RESULTS: In a series of 21 colonoscopies in 20 patients: peak pushing force = 4.4 kg, pulling force = -1.8 kg, anti-clockwise torque = 1.0 Newton meters, clockwise torque = 0.8 Newton meters. Percentage time force greater than 1 kg = 5%. Peak anal insertion force = 1.8 kg.
CONCLUSIONS: These measurements represent the first accurate measurements of the forces exerted during colonoscopy. Reducing the force during colonoscopy is likely to diminish pain and reduce the risk of perforation. A knowledge of these forces may also help with the design of new instruments and models for teaching or research.