The measurement of forces exerted during colonoscopy.

Appleyard M N, Mosse C A, Mills T N, Bell G D, Castillo F D, Swain C P

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The widely varying forces used at colonoscopy have not been measured. An electronic device was designed to measure the forces exerted by the clinician on the endoscope during colonoscopy.

METHODS

The device featured a handle designed in the shape of a hinged split cylinder that could be locked around the endoscope but readily moved up and down the insertion tube as the colonoscopy proceeded. This cylinder contained strain-gauges arranged so that the forces transmitted could be accurately measured. The device recorded the torque forces in addition to the push and pull forces exerted during diagnostic colonoscopy.

RESULTS

In a series of 21 colonoscopies in 20 patients: peak pushing force = 4.4 kg, pulling force = -1.8 kg, anti-clockwise torque = 1.0 Newton meters, clockwise torque = 0.8 Newton meters. Percentage time force greater than 1 kg = 5%. Peak anal insertion force = 1.8 kg.

CONCLUSIONS

These measurements represent the first accurate measurements of the forces exerted during colonoscopy. Reducing the force during colonoscopy is likely to diminish pain and reduce the risk of perforation. A knowledge of these forces may also help with the design of new instruments and models for teaching or research.

Для цитирования: Appleyard M N, Mosse C A, Mills T N, Bell G D, Castillo F D, Swain C P. The measurement of forces exerted during colonoscopy.. Gastrointestinal endoscopy. 2000. DOI: 10.1067/mge.2000.107218.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.