Reliability and validity of a new device to measure isometric strength in polyarticular exercises.

Verdera F, Champavier L, Schmidt C, Bermon S, Marconnet P

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to assess the validity of a new device to evaluate isometric strength during multi-joint exercise such as the squat or bench press.

METHODS

The device used an electric motor-driven bar to simultaneously generate and measure forces during weight lifting exercises. This new device and a force platform measured the forces generated by either the motor on a fixed telescopic steel girder (passive condition) or a subject pushing vertically against the bar from three squat positions (active condition). In the passive condition, 252 measurements were made, with 3 trials for 6 bar heights and 14 bar loads. In the active condition, 8 young physically active students (age, height and body mass were 25.1 +/- 2.6 years, 179.3 +/- 7.2 cm and 82.0 +/- 9.9 kg, respectively) performed 3 maximal isometric strength (MIS) trials in each of the 3 squat exercise positions (parallel, half and quarter squat), and 3 additional MIS trials in one position randomly assigned two weeks later to test inter-day reliability.

RESULTS

In the passive condition, no differences were observed between the forces measured by the force platform and the new device. The coefficient of linear regression (r) and the coefficient of variation (CV) were 1 and below 0.23%, respectively. In the active condition, the peak MIS measured was 2828 N and the values of r and CV were above 0.982 and below 5.96%, respectively. The assessment of inter-day reliability showed an r value of 0.984 and a CV of 3.98%.

CONCLUSIONS

This study demonstrated that the new electric motor-driven exerciser provides valid and reliable data when used to generate forces and measure isometric strength throughout the load and motion ranges commonly used in squat exercise.

Для цитирования: Verdera F, Champavier L, Schmidt C, Bermon S, Marconnet P. Reliability and validity of a new device to measure isometric strength in polyarticular exercises.. The Journal of sports medicine and physical fitness. 1999.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.