DOI 10.1007/s10840-026-02427-2

Медицинская библиотека · Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026

Clinical and experimental evaluation of averaged impedance drop during radiofrequency catheter ablation.

Nakamura Nobuhiro, Oka Satoshi, Wakamiya Akinori, Ueda Nobuhiko, Nakajima Kenzaburo, Kamakura Tsukasa, Wada Mitsuru, Inoue Yuko, Miyamoto Koji, Morita Hideaki, Kusano Kengo

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Real-time monitoring of myocardial tissue response during radiofrequency applications using a mesh-shaped flexible irrigation tip ablation catheter (TactiFlex) remains challenging. This study assessed the dynamics of averaged impedance drop percentage (AID%), a novel ablation parameter, and its association with conduction gaps and lesion characteristics.

METHODS

We retrospectively analyzed 1,423 radiofrequency applications in 30 patients undergoing initial pulmonary vein isolation using TactiFlex to examine the relationship between AID% and radiofrequency applications within conduction gap segments. Mechanistic validation was performed using an ex-vivo porcine myocardial model (152 lesions).

RESULTS

Among 1,423 radiofrequency applications, 138 were located within conduction gap segments. AID% was lower in radiofrequency applications in conduction gap segments than in non-conduction gap segments, although the difference was modest (10.2 [8.3-12.2] % vs. 10.9 [9.0-13.1] %, p = 0.002). The overall discriminative performance was limited (area under the curve, 0.58; 95% confidence interval, 0.53-0.63). In mixed-effects logistic regression accounting for within-patient clustering, AID% <10% was not significantly associated with radiofrequency applications in conduction gap segments (odds ratio, 1.560; 95% confidence interval, 0.970-2.500; p = 0.065), whereas average radiofrequency power was significant. In the ex vivo study, lesion characteristics varied according to radiofrequency power, contact force, and duration, even under fixed target AID% conditions.

CONCLUSIONS

AID% reflects a composite impedance-based tissue response to radiofrequency delivery that is influenced by multiple ablation parameters. Its limited ability to discriminate conduction-gap segments, together with the variability in lesion characteristics across radiofrequency delivery conditions, suggests that AID% should be interpreted as a marker of tissue response rather than as a stand-alone lesion-quality metric. The averaged impedance drop percentage (AID%) reflects a composite impedance-based tissue response to radiofrequency delivery that is influenced by multiple ablation parameters. Its limited ability to discriminate conduction gap segments, together with the variability in lesion characteristics across radiofrequency delivery conditions, suggests that AID% should be interpreted as a marker of tissue response rather than as a stand-alone lesion-quality metric.

Для цитирования: Nakamura Nobuhiro, Oka Satoshi, Wakamiya Akinori, Ueda Nobuhiko, Nakajima Kenzaburo, Kamakura Tsukasa, Wada Mitsuru, Inoue Yuko, Miyamoto Koji, Morita Hideaki, Kusano Kengo. Clinical and experimental evaluation of averaged impedance drop during radiofrequency catheter ablation.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2026. DOI: 10.1007/s10840-026-02427-2.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.