DOI 10.2147/rmhp.s576606

Медицинская библиотека · Risk management and healthcare policy · 2026

Failure Mode and Effects Analysis in Reducing the Incidence of Sterilization Cycle Interruptions in Low-Temperature Hydrogen Peroxide Plasma Sterilizers.

Yi Liangying, You Ling, Pan Wei, Wu Kaili, Zhang Hongxia

СтатьяПеревод в очереди

Аннотация

BACKGROUND

This study aimed to investigate the efficacy of failure mode and effects analysis (FMEA) in optimizing the sterilization process for low-temperature hydrogen peroxide plasma sterilizers.

METHODS

Purposive sampling was used to select low-temperature hydrogen peroxide plasma sterilizers from a hospital in China as the study objects. The operational data for 1551 sterilization cycles in the low-temperature hydrogen peroxide plasma sterilizers, from May 2024 to November 2024, were classified into the control group. The operational data for 1492 sterilization cycles in the low-temperature hydrogen peroxide plasma sterilizers, from January 2025 to June 2025 following the implementation of the FMEA-based risk management scheme, were classified into the experimental group. The incidence of sterilization cycle interruptions, the risk priority number scores for high-risk failure modes, and the incidence of high-risk failure modes in both groups were investigated.

RESULTS

After the optimization of the operating procedures, the incidence of sterilization cycle interruptions decreased from 5.48% (85/1551) to 0.87% (13/1492); the risk priority number scores and the incidence of high-risk failure modes were also decreased (P < 0.05).

CONCLUSION

The FMEA-based risk management scheme could optimize the sterilization process for low-temperature hydrogen peroxide plasma sterilizers and ensure the quality of medical device sterilization.

Для цитирования: Yi Liangying, You Ling, Pan Wei, Wu Kaili, Zhang Hongxia. Failure Mode and Effects Analysis in Reducing the Incidence of Sterilization Cycle Interruptions in Low-Temperature Hydrogen Peroxide Plasma Sterilizers.. Risk management and healthcare policy. 2026. DOI: 10.2147/rmhp.s576606.