Surface finish-dependent stability of stainless steel under long-term cleaning, disinfection and sterilization process: AISI 420 in medical applications.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Barmouz M, Steinhäuser F, Azarhoushang B, Bucher V, Sauer D
Аннотация
BACKGROUND: Reusable stainless-steel medical devices are subjected to repeated cleaning, disinfection, and sterilization processes that may progressively alter surface properties, potentially affecting durability, and clinical performance.
METHODS: The long-term effects of up to 1000 processing cycles were investigated on AISI 420 (American Iron and Steel Institute grade 420) stainless steel samples with two surface preparations: mechanically brushed (MB) and glass bead -blasted (GBB). Surface roughness (Ra, Sa, Sq), wettability (contact angle), microhardness, and surface morphology were evaluated before and after processing using profilometry, contact angle measurements, microhardness testing, and SEM/EDX analysis.
RESULTS: GBB surfaces consistently exhibited higher roughness (Ra ≈ 0.58 µm) than MB surfaces (Ra ≈ 0.42-0.46 µm) and remained more hydrophilic, with significantly lower contact angles (≈ 40-50°) compared to MB samples (≈ 65-70°). Microhardness values were higher for GBB samples (≈ 820-870 HV0.05) than for MB samples (≈ 680-705 HV0.05) across all processing cycles. However, SEM/EDX analysis revealed more frequent oxidation-related dark spots on GBB surfaces after extended cycling, indicating increased susceptibility to surface discoloration.
CONCLUSIONS: Surface preparation has a significant influence on the long-term stability and performance of reusable stainless-steel medical devices. While glass bead-blasted surfaces offer higher hardness and wettability, they may be more prone to oxidation-related discoloration during prolonged reprocessing, with potential implications for infection prevention and device longevity.
METHODS: The long-term effects of up to 1000 processing cycles were investigated on AISI 420 (American Iron and Steel Institute grade 420) stainless steel samples with two surface preparations: mechanically brushed (MB) and glass bead -blasted (GBB). Surface roughness (Ra, Sa, Sq), wettability (contact angle), microhardness, and surface morphology were evaluated before and after processing using profilometry, contact angle measurements, microhardness testing, and SEM/EDX analysis.
RESULTS: GBB surfaces consistently exhibited higher roughness (Ra ≈ 0.58 µm) than MB surfaces (Ra ≈ 0.42-0.46 µm) and remained more hydrophilic, with significantly lower contact angles (≈ 40-50°) compared to MB samples (≈ 65-70°). Microhardness values were higher for GBB samples (≈ 820-870 HV0.05) than for MB samples (≈ 680-705 HV0.05) across all processing cycles. However, SEM/EDX analysis revealed more frequent oxidation-related dark spots on GBB surfaces after extended cycling, indicating increased susceptibility to surface discoloration.
CONCLUSIONS: Surface preparation has a significant influence on the long-term stability and performance of reusable stainless-steel medical devices. While glass bead-blasted surfaces offer higher hardness and wettability, they may be more prone to oxidation-related discoloration during prolonged reprocessing, with potential implications for infection prevention and device longevity.