Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Fernández-Rodríguez Diana, Tarabichi Saad, Golankiewicz Krystal, Zappley Nicolina, Parvizi Javad
Аннотация
BACKGROUND: One important factor for the prevention of surgical site infections is ultraclean air in the operating room (OR). Still, the direct sterilization potential of most technologies, especially in a dynamic clinical setting, is not well understood. We aimed to determine and compare the microbial presence from the inlet and outlet flow of a filtration unit with crystalline ultraviolet-C (C-UVC) light.
METHODS: A prospective study was conducted at a single institution, where primary total joint arthroplasty and spine surgeries were performed. The OR was fitted with a positive ventilation system. In addition, a filtration unit with a C-UVC sterilizing light was placed in the OR. The inlet and outlet air flows were swabbed simultaneously and compared. Swabs were processed for culture and next-generation sequencing.
RESULTS: The mean length of the surgical procedures sampled was 68 ± 13 minutes. Overall, 19 out of 200 (9.5%) swabs isolated microorganisms. Inlet air swabs were positive at a higher rate (16 versus 3%; P < .01) compared to the outlet air swabs. A wide variety of Gram-positive, Gram-negative, and anaerobic bacteria were isolated, but fungi were only recovered from inlet air swabs. The detection of microorganisms was also higher when more door openings were performed (32.5 ± 7.1 versus 27.9 ± 5.6; P < .01).
CONCLUSIONS: Air swabs mainly isolated microorganisms from the inlet flow to the filtration unit with a C-UVC light. The sterilizing unit counteracted factors affecting the air quality in the OR, namely door openings, surgical personnel, and tissue combustion.
METHODS: A prospective study was conducted at a single institution, where primary total joint arthroplasty and spine surgeries were performed. The OR was fitted with a positive ventilation system. In addition, a filtration unit with a C-UVC sterilizing light was placed in the OR. The inlet and outlet air flows were swabbed simultaneously and compared. Swabs were processed for culture and next-generation sequencing.
RESULTS: The mean length of the surgical procedures sampled was 68 ± 13 minutes. Overall, 19 out of 200 (9.5%) swabs isolated microorganisms. Inlet air swabs were positive at a higher rate (16 versus 3%; P < .01) compared to the outlet air swabs. A wide variety of Gram-positive, Gram-negative, and anaerobic bacteria were isolated, but fungi were only recovered from inlet air swabs. The detection of microorganisms was also higher when more door openings were performed (32.5 ± 7.1 versus 27.9 ± 5.6; P < .01).
CONCLUSIONS: Air swabs mainly isolated microorganisms from the inlet flow to the filtration unit with a C-UVC light. The sterilizing unit counteracted factors affecting the air quality in the OR, namely door openings, surgical personnel, and tissue combustion.