Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit.

Fernández-Rodríguez Diana, Tarabichi Saad, Golankiewicz Krystal, Zappley Nicolina, Parvizi Javad

СтатьяПеревод в очереди

Аннотация

BACKGROUND

One important factor for the prevention of surgical site infections is ultraclean air in the operating room (OR). Still, the direct sterilization potential of most technologies, especially in a dynamic clinical setting, is not well understood. We aimed to determine and compare the microbial presence from the inlet and outlet flow of a filtration unit with crystalline ultraviolet-C (C-UVC) light.

METHODS

A prospective study was conducted at a single institution, where primary total joint arthroplasty and spine surgeries were performed. The OR was fitted with a positive ventilation system. In addition, a filtration unit with a C-UVC sterilizing light was placed in the OR. The inlet and outlet air flows were swabbed simultaneously and compared. Swabs were processed for culture and next-generation sequencing.

RESULTS

The mean length of the surgical procedures sampled was 68 ± 13 minutes. Overall, 19 out of 200 (9.5%) swabs isolated microorganisms. Inlet air swabs were positive at a higher rate (16 versus 3%; P < .01) compared to the outlet air swabs. A wide variety of Gram-positive, Gram-negative, and anaerobic bacteria were isolated, but fungi were only recovered from inlet air swabs. The detection of microorganisms was also higher when more door openings were performed (32.5 ± 7.1 versus 27.9 ± 5.6; P < .01).

CONCLUSIONS

Air swabs mainly isolated microorganisms from the inlet flow to the filtration unit with a C-UVC light. The sterilizing unit counteracted factors affecting the air quality in the OR, namely door openings, surgical personnel, and tissue combustion.

Для цитирования: Fernández-Rodríguez Diana, Tarabichi Saad, Golankiewicz Krystal, Zappley Nicolina, Parvizi Javad. Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit.. The Journal of arthroplasty. 2024. DOI: 10.1016/j.arth.2024.05.072.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.