DOI 10.1371/journal.pone.0284967

Медицинская библиотека · PloS one · 2023

Adenosine triphosphate (ATP) sampling algorithm for monitoring the cleanliness of surgical instruments.

Pontes Daniela Oliveira, Costa Dayane de Melo, da Silva Pereira Priscilla Perez, Whiteley Greg S, Glasbey Trevor, Tipple Anaclara Ferreira Veiga

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Timely detection of cleaning failure is critical for quality assurance within Sterilising Service Units (SSUs). Rapid Adenosine Triphosphate (ATP) testing provides a real time and quantitative indication of cellular contaminants, when used to measure surface or device cleanliness. The aim of this study was to investigate the use of an ATP algorithm and to whether it could be used as a routine quality assurance step, to monitor surgical instruments cleanliness in SSUs prior to sterilisation.

METHODS

Cleanliness monitoring using rapid ATP testing was undertaken in the SSUs of four hospitals located in the western (Amazonia) region of Brazil. ATP testing was conducted (Clean Trace, 3M) on 163 surgical instruments, following manual cleaning. A sampling algorithm using a duplicate swab approach was applied to indicate surgical instruments as (i) very clean, (ii) clean, (iii) equivocal or (iv) fail, based around a 'clean' cut-off of 250 Relative Light Units (RLU) and a 'very clean' <100 RLU.

RESULTS

The four cleanliness categories were significantly differentiated (P≤0.001). The worst performing locations (hospitals A & C) had failure rates of 39.2% and 32.4%, respectively, and were distinctly different from hospitals B & D (P≤0.001). The best performing hospitals (B & D) had failure rates of 7.7% and 2.8%, respectively.

CONCLUSION

The ATP testing algorithm provides a simple to use method within SSUs. The measurements are in real time, quantitative and useful for risk-based quality assurance monitoring, and the tool can be used for staff training. The four-tiered approach to the grading of surgical instrument cleanliness provides a nuanced approach for continuous quality improvement within SSU than does a simple pass/fail methodology.

Для цитирования: Pontes Daniela Oliveira, Costa Dayane de Melo, da Silva Pereira Priscilla Perez, Whiteley Greg S, Glasbey Trevor, Tipple Anaclara Ferreira Veiga. Adenosine triphosphate (ATP) sampling algorithm for monitoring the cleanliness of surgical instruments.. PloS one. 2023. DOI: 10.1371/journal.pone.0284967.