m+Medelevator.RU
← Поиск публикаций
European journal of dentistry · 2023

Determination of the Sterilization Dose of Gamma-Ray Irradiation for Polyvinyl Alcohol-Collagen-Chitosan Composite Membrane as a Material for Periodontal Regenerative Surgery.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Susanto Agus, Komara Ira, Beatrix Maria Theresia, Lukitowati Fajar, Amaliya Amaliya, Hendiani Ina, Miranda Aldilla

Аннотация

OBJECTIVE: Membrane sterility is very necessary considering its function as an implant material. Therefore, this research aims to determine the dose of gamma-ray irradiation for the sterilization of polyvinyl alcohol (PVA)-collagen-chitosan composite membranes used as regenerative surgery materials.

MATERIALS AND METHODS: A total of 100 pieces of the composite membranes were prepared in a size of 2.0 × 1.5 cm by mixing 7.5% PVA, 3% collagen, and 2% chitosan using the film casting method in three batches. Furthermore, the bioburden test was performed to determine the initial microbial count in the sample by following ISO 11737-1. The results were used to ascertain the dose of gamma-ray irradiation on the sample according to ISO 11137-2. The dose verification test was then performed at the sterility assurance level 10-6.

RESULTS: The average result of the bioburden test from three batches was 6.6 colony forming unit; hence, the verification dose was 4.8 kGy. In the verification dose test, since there was only one contaminated sample, the sterility dose test was continued.

CONCLUSION: The sterile gamma-ray irradiation dose for PVA-collagen-chitosan composite membrane was 17.1 kGy.

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты