American journal of infection control · 2023
How effective are the alcohol flush and drying cycles of automated endoscope reprocessors? Stripped endoscope model.
Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yassin Mohamed, Clifford Adrian, Dixon Heather, Donskey Curtis J
Аннотация
BACKGROUND: Effective drying of the internal channels of endoscopes is essential to prevent the growth of water-borne pathogens and to assure adequate sterilization with vaporized hydrogen peroxide or ethylene oxide. The aim of this study was to evaluate the dryness of endoscopes after a routine disinfection process in an automated endoscope reprocessor.
METHODS: Stripped endoscopes (SE) that allow for visual inspection of the inside channels were reprocessed per protocol in a large urban medical center, with a 3-minute or 10-minute air flush following reprocessing. SE was hung and observed for any water within the channels after reprocessing and after a week of ambient storage. Ready-for-use endoscopes were also randomly spot-checked for moisture visually and with moisture detection paper.
RESULTS: All SE were grossly wet after HLD with a 3-minute air flush, despite alcohol flush and drying cycle. The 10-minute air flush was effective at drying the biopsy/suction channel, but not the air/water channels. Hanging had limited effect, being most effective in the biopsy/suction channels. Of the 77 ready-for-use respiratory and gastrointestinal endoscopes assessed, 37 (48.1%) showed evidence of retained moisture.
CONCLUSIONS: Air flush cycles commonly used in the final steps of automated endoscope reprocessing may not adequately dry endoscope channels, particularly the narrower diameter air/water channels. An extended 10-minute air flush appears effective at drying the larger biopsy/suction channel, but has limited effect on the air/water channels.
METHODS: Stripped endoscopes (SE) that allow for visual inspection of the inside channels were reprocessed per protocol in a large urban medical center, with a 3-minute or 10-minute air flush following reprocessing. SE was hung and observed for any water within the channels after reprocessing and after a week of ambient storage. Ready-for-use endoscopes were also randomly spot-checked for moisture visually and with moisture detection paper.
RESULTS: All SE were grossly wet after HLD with a 3-minute air flush, despite alcohol flush and drying cycle. The 10-minute air flush was effective at drying the biopsy/suction channel, but not the air/water channels. Hanging had limited effect, being most effective in the biopsy/suction channels. Of the 77 ready-for-use respiratory and gastrointestinal endoscopes assessed, 37 (48.1%) showed evidence of retained moisture.
CONCLUSIONS: Air flush cycles commonly used in the final steps of automated endoscope reprocessing may not adequately dry endoscope channels, particularly the narrower diameter air/water channels. An extended 10-minute air flush appears effective at drying the larger biopsy/suction channel, but has limited effect on the air/water channels.