[Qualification of low temperature vaporized hydrogen peroxide sterilization: Towards greater harmonization of practices?].
Аннотация
OBJECTIVES
Comparing performance qualification procedures for low temperature vaporized hydrogen peroxide sterilization. Assessing conformity with draft standard ISO/DIS 22441.
METHODS
Qualification reports from several providers have been compared according to specific criteria: choices of cycles, loads, sterile barrier systems, probes, biological and chemical indicators; checking of packaging integrity and exposure to sterilizing agent.
RESULTS
Six out of 8 reports based on 4 distinct sterilizers have been performed by third-party providers. Routine and process challenge devices are respectively used in 6 and 3 of these reports. Sizes and masses are never mentioned whereas load configuration is always specified. All reports use at least one biological indicator and 50% of them use one chemical indicator at a minimum. Most frequent wrapping materials are Ultra® and Tyvek® bags (respectively 50% and 37.5% of reports). Each qualification monitors per process pression and temperature, and 37.5% of them also quantify hydrogen peroxide concentration. Packaging integrity and environmental exposure are checked in respectively 50% and 12.5% of all reports. All reports have received providers approval.
CONCLUSION
Qualification procedure is based on steam sterilization NF EN 14937 standard, which seems unsuitable for low temperature process. The lack of autonomy, the heterogeneity of loads and measurement choices reveal a low harmonization of practices. New standard should dispel the doubts about this heterogeneity.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.