Splash generation and droplet dispersal in a well-designed, centralized high-level disinfection unit.

Ofstead Cori L, Hopkins Krystina M, Daniels Frank E, Smart Abigail G, Wetzler Harry P

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Sterile processing personnel routinely decontaminate medical devices that are heavily soiled with blood, tissue, and secretions. Contamination may spread throughout processing areas, potentially exposing personnel and patient-ready devices, especially when there is insufficient separation between the dirty and clean areas.

OBJECTIVE

This study aimed to identify activities that generate splash, determine how far droplets travel during manual cleaning, characterize the impact of practices on splash generation, and assess effectiveness of personal protective equipment (PPE) at preventing splash exposure to technicians and visitors in the decontamination unit.

METHODS

Moisture-detection paper was affixed to PPE and environmental surfaces in a new processing department designed to optimize workflow and prevent cross-contamination. Droplet generation and dispersal were assessed during manual cleaning of a colonoscope and a transvaginal ultrasound probe.

RESULTS

Splash was generated by most activities and droplets were detected up to 7.25 feet away. Transporting wet endoscopes dispersed droplets on a 15-foot path from the sink to the automated endoscope reprocessor. Extensive droplets were detected on PPE worn by technicians at the sink and observers 3-4 feet away.

CONCLUSIONS

Manual cleaning of devices generated substantial splash, drenching technicians and the environment with droplets that traveled more than 7 feet. Engineering controls and better PPE are needed to reduce personnel exposure and risks associated with the potential dispersal of contaminated fluids throughout the facility.

Для цитирования: Ofstead Cori L, Hopkins Krystina M, Daniels Frank E, Smart Abigail G, Wetzler Harry P. Splash generation and droplet dispersal in a well-designed, centralized high-level disinfection unit.. American journal of infection control. 2022. DOI: 10.1016/j.ajic.2022.08.016.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.