MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jhin.2022.08.010

Медицинская библиотека · The Journal of hospital infection · 2022

Prevalence of bacterial burden on macroscopic contaminants of orthopaedic surgical instruments following sterilization.

Wellington I J, Schneider T J, Hawthorne B C, McCarthy M B, Stelzer J W, Connors J P, Dorsey C, Williams V, Lindsay A, Solovyova O

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Macroscopic contamination of orthopaedic instruments with particulates, including cortical bone and polymethyl methacrylate (PMMA) bone cement, that have previously undergone pre-operative sterilization is frequently encountered peri- or intraoperatively, calling into question the sterility of such instruments.

AIM: To determine if macroscopic contaminants of orthopaedic surgical instrumentation maintain a bacterial burden following sterile processing, and to determine the most commonly contaminated instruments and the most common contaminants.

METHODS

Macroscopic contaminants in orthopaedic instrument trays were collected prospectively at a single tertiary referral centre over a 6-month period from August 2021 to May 2022. When identified, these specimens were swabbed and plated on sheep blood agar. All specimens were incubated at 37 °C for 14 days, and inspected visually for colony formation. When bacterial colony formation was identified, samples were sent for species identification.

RESULTS

In total, 33 contaminants were tested, and only one contaminant was found to be growing bacterial colonies (Corynebacterium sp.). The items most commonly found to have macroscopic contamination were surgical trays (N=9) and cannulated drills (N=7). The identifiable contaminants were bone (N=10), PMMA bone cement (N=4) and hair (N=4). Eleven macroscopic contaminants were not identifiable.

CONCLUSION

This study found that 97% of macroscopic orthopaedic surgical instrument contaminants that underwent sterile processing did not possess a bacterial burden. Contaminants discovered during a procedure are likely to be sterile, and do not pose a substantially increased risk of infection to a patient.

Для цитирования: Wellington I J, Schneider T J, Hawthorne B C, McCarthy M B, Stelzer J W, Connors J P, Dorsey C, Williams V, Lindsay A, Solovyova O. Prevalence of bacterial burden on macroscopic contaminants of orthopaedic surgical instruments following sterilization.. The Journal of hospital infection. 2022. DOI: 10.1016/j.jhin.2022.08.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.