Surgical helmets can be converted into efficient disinfectable powered air-purifying respirators.

Temmesfeld Max Joachim, Gorzkowska-Sobas Agnieszka Anna, Hedlund Kenny, Øyen Martin Øhlund, Kanten Lars, Grant Peter, Jakobsen Rune B

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Filtering facepiece respirators often fail to provide sufficient protection due to a poor fit. Powered air-purifying respirators (PAPRs) are not designed for healthcare personnel, and are challenging to disinfect. Surgical helmets (SH) are available in many United States hospitals but do not provide respiratory protection. Several modifications to SH have been suggested, but none are sufficiently compliant with safety and efficiency standards. The purpose of this investigation was the development of a filter adaptor, which converts SHs into efficient, safe, and disinfectable PAPRs.

METHODS

Four critical features were investigated close to regulatory requirements: total inward leakage of particles, CO2 concentrations, intra-helmet differential pressure, and automated disinfection.

RESULTS

The average total inward leakage in the 2 independent tests were 0.005% and 0.01%. CO2 concentrations were lower than in the original SH. The modification generates a positive differential pressure. The filter's performance was not compromised after 50 cycles in a sterilization machine.

DISCUSSION: The modified SH provides several hundred times better protection than FFP-3 masks.

CONCLUSIONS

Surgical helmets can be modified into safe, efficient, and disinfectable PAPRs, suitable for HCP and the operating room in particular. They can play a role in the preparedness for upcoming events requiring efficient respiratory protection.

Для цитирования: Temmesfeld Max Joachim, Gorzkowska-Sobas Agnieszka Anna, Hedlund Kenny, Øyen Martin Øhlund, Kanten Lars, Grant Peter, Jakobsen Rune B. Surgical helmets can be converted into efficient disinfectable powered air-purifying respirators.. American journal of infection control. 2021. DOI: 10.1016/j.ajic.2021.12.002.