OBJECTIVES

The purpose of this study was to evaluate the dimensional stability of prototyped surgical guides after autoclave and 2% glutaraldehyde sterilization processes.

MATERIAL AND

METHODS

Twenty prototyped surgical guides were prepared and submitted to two sterilization processes (n = 10): Physics - sterilization by autoclave (saturated water vapor under pressure, temperature of 126 to 130 °C, pressure of 1,7 at 1.9 kgf/cm2, 16 min); Chemistry - sterilization by 2% glutaraldehyde for 10 h. Six pre-established points were measured in the prototyped surgical guides, before and after sterilization, using a digital caliper rule. The comparisons were made using orthogonal contrasts using the linear model of mixed effects (random and fixed).

RESULTS

there are no significant differences between autoclave and glutaraldehyde 2% (p>0.05) there are significant differences after autoclave sterilization (p<0.05) and there are not significant differences after 2% glutaraldehyde sterilization (p>0.05).

CONCLUSION

autoclave promoted dimensional alteration of the prototyped surgical guides, and the chemical sterilization by glutaraldehyde 2% did not cause dimensional alteration of the prototype surgical guides, being a favorable choice for sterilization.

CLINICAL SIGNIFICANCE: the sterilization of surgical guides can be performed through the chemical process with 2% glutaraldehyde without changing the linear precision of the prototype surgical guides.

Для цитирования: Matheus Diego Moraes De, Gomes Erica Alves, Barcellos Daphne Camara, Rosetti Elizabeth Pimentel, Margonar Rogério. Linear dimensional accuracy of stereolithographic surgical guide after chemistry and physics sterilization.. Journal of stomatology, oral and maxillofacial surgery. 2021. DOI: 10.1016/j.jormas.2021.11.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.