Archivio italiano di urologia, andrologia : organo ufficiale [di] Societa italiana di ecografia urologica e nefrologica · 2021
Comparison of a single-use, digital flexible ureteroscope with a reusable, fiberoptic ureteroscope for management of patients with urolithiasis.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Mourmouris Panagiotis, Tzelves Lazaros, Raptidis Grigorios, Berdempes Marinos, Markopoulos Titos, Dellis Grigorios, Siafakas Ioannis, Skolarikos Andreas
Аннотация
OBJECTIVES: Ureteroscopy is one of the commonest procedures performed to manage urolithiasis. Flexible ureteroscopy has been traditionally based on reusable, fiber-optic ureteroscopes. Technology advancements permitted the development of single-use scopes with digital image. The aim of this study is to compare efficacy and safety between a reusable, fiberoptic ureteroscope with a single-use, digital scope.
PATIENTS AND METHODS: We collected data based on chart review from a prospectively collected database on a tertiary, high-volume hospital in Greece. Baseline, perioperative and postoperative data were gathered and analyzed. Chi-square and Fisher's exact test was used to compare qualitative data and unpaired t-test for continuous data, with a statistical significance set at a = 0.05.
RESULTS: 40 patients underwent flexible ureteroscopy with a single- use digital scope, while 37 with the reusable scope. The two groups were matched regarding baseline characteristics and stone-related parameters. After data analysis, a shorter operative time in favor of single-use flexible ureteroscope was detected (45 vs 65 min, p = 0.001), while safety was also in favor of this type of scope with a significantly higher immediate stonefree rate (70% vs 43%, p = 0.005). Overall complications did not differ between the two groups, although a lower sepsis rate was detected in patients treated with single-use scope.
CONCLUSIONS: Our findings indicate that single-use, digital ureteroscopes are a viable alternative for flexible ureteroscopy and management of urolithiasis, especially in centers with deficient facilities for sterilization and ensured funds for more expensive reusable scopes.
PATIENTS AND METHODS: We collected data based on chart review from a prospectively collected database on a tertiary, high-volume hospital in Greece. Baseline, perioperative and postoperative data were gathered and analyzed. Chi-square and Fisher's exact test was used to compare qualitative data and unpaired t-test for continuous data, with a statistical significance set at a = 0.05.
RESULTS: 40 patients underwent flexible ureteroscopy with a single- use digital scope, while 37 with the reusable scope. The two groups were matched regarding baseline characteristics and stone-related parameters. After data analysis, a shorter operative time in favor of single-use flexible ureteroscope was detected (45 vs 65 min, p = 0.001), while safety was also in favor of this type of scope with a significantly higher immediate stonefree rate (70% vs 43%, p = 0.005). Overall complications did not differ between the two groups, although a lower sepsis rate was detected in patients treated with single-use scope.
CONCLUSIONS: Our findings indicate that single-use, digital ureteroscopes are a viable alternative for flexible ureteroscopy and management of urolithiasis, especially in centers with deficient facilities for sterilization and ensured funds for more expensive reusable scopes.