DOI 10.1097/md.0000000000027492

Медицинская библиотека · Medicine · 2021

The packaging and clean method contribute to insulation failure of electrosurgical instruments.

Zhang Ying, Zhang Yanyan, Wang Yafei, Yang Lili, Hu Ruying

СтатьяПеревод в очереди

Аннотация

With the rapid development of medical technology, the use of electrosurgical instruments is dramatically increased in various types of surgery. However, the damage of the insulation layer of the reusable electrosurgical instrument often causes surgical accidents. The procedures of packaging and cleaning contribute to many damages to insulating layer of reusable electrosurgical instruments.Various types of reusable electrosurgical instruments were detected for insulation failures, conduction failures, short-circuit by using a high-voltage detector, DIATEG (Morgate company). In addition, reusable electrosurgical instruments were detected for insulation failures after packaging and cleaning by different procedures.13.1% (129/740) electrosurgical instruments had an insulation test failure; 6.2% (9/146) monopolar wires were with conduction failure; and 7.7% (16/207) bipolar wires were with short-circuit. Different packaging and cleaning procedures contribute to various degrees of damages to insulating property of reusable electrosurgical instruments.Insulation failure was a wide problem of reusable electrosurgical instruments, while fixed packaging method and mild cleaning procedures result in fewer damages to insulating property of reusable electrosurgical instruments.

Для цитирования: Zhang Ying, Zhang Yanyan, Wang Yafei, Yang Lili, Hu Ruying. The packaging and clean method contribute to insulation failure of electrosurgical instruments.. Medicine. 2021. DOI: 10.1097/md.0000000000027492.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.