Monitoring the Effective Sterilization of Low-Temperature Hydrogen Peroxide Gas Plasma Sterilizers in 58 Hospitals - 22 PLADs, China, June 2015-December 2019.

Wang Jiaqi, Zhang Baoying, Sun Huihui, Zhang Jian, Duan Hongyang, Ban Haiqun, Shen Jin, Zhang Liubo

СтатьяПеревод в очереди

Аннотация

WHAT IS ALREADY KNOWN ON THIS TOPIC?: Hydrogen peroxide sterilizeation is widely used for luminal devices. However, the low penetrability of the sterilant is of major concern.

WHAT IS ADDED BY THIS REPORT?: This report investigated the effective sterilization of low-temperature hydrogen peroxide gas plasma sterilizers and compared the applicability of different biological monitoring methods based on medical luminal devices.

WHAT ARE THE IMPLICATIONS FOR PUBLIC HEALTH PRACTICE?: It is recommended to use a biological process challenge device for monitoring the sterilization of luminal devices with low-temperature hydrogen peroxide gas plasma sterilizers.

Для цитирования: Wang Jiaqi, Zhang Baoying, Sun Huihui, Zhang Jian, Duan Hongyang, Ban Haiqun, Shen Jin, Zhang Liubo. Monitoring the Effective Sterilization of Low-Temperature Hydrogen Peroxide Gas Plasma Sterilizers in 58 Hospitals - 22 PLADs, China, June 2015-December 2019.. China CDC weekly. 2021. DOI: 10.46234/ccdcw2021.152.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.