Does blood on "dirty" instruments interfere with the effectiveness of sterilization technologies?

Rutala William A, Gergen Maria F, Weber David J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

We evaluated the robustness of sterilization technologies when spores and bacteria were placed on "dirty" instruments and overlaid with blood. The results illustrate that steam sterilization is the most effective sterilization technology with the largest margin of safety, followed by ethylene oxide and hydrogen peroxide gas plasma.

Для цитирования: Rutala William A, Gergen Maria F, Weber David J. Does blood on "dirty" instruments interfere with the effectiveness of sterilization technologies?. Infection control and hospital epidemiology. 2021. DOI: 10.1017/ice.2021.202.