MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ajic.2020.08.034

Медицинская библиотека · American journal of infection control · 2020

Efficacy of flexible endoscope drying using novel endoscope test articles that allow direct visualization of the internal channel systems.

Nerandzic Michelle, Antloga Kathleen, Litto Christine, Robinson Nancy

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Thorough drying of flexible endoscope channels has been identified as an essential reprocessing step. Yet, instructions are not specific on how to dry endoscopes. There is lack of data supporting efficacy of current drying practices, due to limitations in determining channel dryness.

METHODS

Novel endoscope test articles were used to evaluate the effectiveness of alcohol flush and hanging in an ambient endoscope storage cabinet. Prepared test articles were hung in a storage cabinet for 5 days and visually inspected for residual liquid. The procedure for preoperative inspection of endoscopic systems was performed to determine the procedure's efficacy for removing residual liquid. Then, testing was performed to assess the impact of pressure, residual liquid type and route of air application on time to dry using compressed air.

RESULTS

Alcohol flush followed by hanging in an ambient storage cabinet was not effective for drying endoscope channels, and residual liquid was not completely removed after performing the steps of the preoperative inspection of endoscopic channels. The factors impacting effective compressed air drying were channel dependent. For some channels, alcohol increased the time to dry.

CONCLUSIONS

Endoscope drying is complex; borescope evaluation does not ensure a dry device.

Для цитирования: Nerandzic Michelle, Antloga Kathleen, Litto Christine, Robinson Nancy. Efficacy of flexible endoscope drying using novel endoscope test articles that allow direct visualization of the internal channel systems.. American journal of infection control. 2020. DOI: 10.1016/j.ajic.2020.08.034.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.