DOI 10.28920/dhm54.3.212-216

Медицинская библиотека · Diving and hyperbaric medicine · 2024

Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation.

Wang Cong, Yu Qiuhong, Liu Yaling, Ren Ziqi, Liu Ying, Xue Lianbi

СтатьяОценочное исследованиеКлинический случайПеревод в очереди

Аннотация

INTRODUCTION

The stability of a new hyperbaric ventilator (Shangrila590, Beijing Aeonmed Company, Beijing, China) at different clinically relevant pressures in a hyperbaric chamber during pressure-controlled ventilation (PCV) was investigated.

METHODS

The ventilator was connected to a test lung in the multiplace hyperbaric chamber. The inspiratory pressure (PI) of the ventilator was set to 1.0, 1.5, 2.0, 2.5 and 3.0 kPa (approximately 10, 15, 20, 25 and 30 cmH₂O). The compliance and resistance of the test lung were set to 200 mL·kPa⁻¹ and 2 kPa·L⁻¹·s⁻¹, respectively. Experiments were conducted at 101, 203 and 284 kPa ambient pressure (1.0, 2.0 and 2.8 atmospheres absolute respectively). At each of the 5 PI values, the tidal volume (VT), peak inspiratory pressure (Ppeak) and peak inspiratory flow (Fpeak) displayed by the ventilator and the test lung were recorded for 20 cycles. Test lung data were considered the actual ventilation values. The ventilation data were compared among the three groups to evaluate the stability of the ventilator.

RESULTS

At every PI, the Ppeak detected by the ventilator decreased slightly with increasing ambient pressure. The Fpeak values measured by the test lung decreased substantially as the ambient pressure increased. Nevertheless, the reduction in VT at 284 kPa and PI 30 cmH₂O (compared to performance at 101 kPa) was comparatively small (approximately 60 ml).

CONCLUSIONS

In PCV mode this ventilator provided relatively stable VT across clinically relevant PI values to ambient pressures as high as 284 kPa. However, because Fpeak decreases at higher ambient pressure, some user adjustment might be necessary for precise VT maintenance during clinical use at higher PIs and ambient pressures.

Для цитирования: Wang Cong, Yu Qiuhong, Liu Yaling, Ren Ziqi, Liu Ying, Xue Lianbi. Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation.. Diving and hyperbaric medicine. 2024. DOI: 10.28920/dhm54.3.212-216.