False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Laranjeira Paulo Roberto, Bronzatti Jeane Aparecida Gonzalez, Bruna Camila Quartim de Moraes, de Souza Rafael Queiroz, Graziano Kazuko Uchikawa, Lusignan Viktoriya
Аннотация
OBJECTIVE: To determine if the standardized B&D type test for hospital steam sterilizer is correctly indicating cycle failures in slower come-up ramps cycles.
METHODS: Two commercially available B&D type test were challenged in a quality control sterilizer. A common failure was simulated in triplicate cycles, using a standardized cycle configuration, and then compared to triplicate cycles of a common cycle configuration. Tests procedures were conducted according to B&D manufacturer test standard and results were compared to standardized endpoint specifications.
RESULTS: We found that B&D type tests are only capable of detecting the presence of non-condensable gases if the sterilization equipment is adjusted to meet all the cycle requirements of the test. False positive results were obtained with come-up ramp time of 3 min. Correct results were only obtained with come-up ramp of 1.7-1.9 min.
CONCLUSIONS: Until the ISO 17665 and AAMI ST-79 standards are revised, equipment qualification experts should observe come-up time duration criterion for B&D type test cycles according to ISO 11140-4. Sterile Processing Department professionals must add the come-up ramp criterion to cycle evaluation before clearing the equipment for routine use. This will allow B&D correct performance, reducing the infection risk from unsterilized medical device.
METHODS: Two commercially available B&D type test were challenged in a quality control sterilizer. A common failure was simulated in triplicate cycles, using a standardized cycle configuration, and then compared to triplicate cycles of a common cycle configuration. Tests procedures were conducted according to B&D manufacturer test standard and results were compared to standardized endpoint specifications.
RESULTS: We found that B&D type tests are only capable of detecting the presence of non-condensable gases if the sterilization equipment is adjusted to meet all the cycle requirements of the test. False positive results were obtained with come-up ramp time of 3 min. Correct results were only obtained with come-up ramp of 1.7-1.9 min.
CONCLUSIONS: Until the ISO 17665 and AAMI ST-79 standards are revised, equipment qualification experts should observe come-up time duration criterion for B&D type test cycles according to ISO 11140-4. Sterile Processing Department professionals must add the come-up ramp criterion to cycle evaluation before clearing the equipment for routine use. This will allow B&D correct performance, reducing the infection risk from unsterilized medical device.