False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study.

Laranjeira Paulo Roberto, Bronzatti Jeane Aparecida Gonzalez, Bruna Camila Quartim de Moraes, de Souza Rafael Queiroz, Graziano Kazuko Uchikawa, Lusignan Viktoriya

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To determine if the standardized B&D type test for hospital steam sterilizer is correctly indicating cycle failures in slower come-up ramps cycles.

METHODS

Two commercially available B&D type test were challenged in a quality control sterilizer. A common failure was simulated in triplicate cycles, using a standardized cycle configuration, and then compared to triplicate cycles of a common cycle configuration. Tests procedures were conducted according to B&D manufacturer test standard and results were compared to standardized endpoint specifications.

RESULTS

We found that B&D type tests are only capable of detecting the presence of non-condensable gases if the sterilization equipment is adjusted to meet all the cycle requirements of the test. False positive results were obtained with come-up ramp time of 3 min. Correct results were only obtained with come-up ramp of 1.7-1.9 min.

CONCLUSIONS

Until the ISO 17665 and AAMI ST-79 standards are revised, equipment qualification experts should observe come-up time duration criterion for B&D type test cycles according to ISO 11140-4. Sterile Processing Department professionals must add the come-up ramp criterion to cycle evaluation before clearing the equipment for routine use. This will allow B&D correct performance, reducing the infection risk from unsterilized medical device.

Для цитирования: Laranjeira Paulo Roberto, Bronzatti Jeane Aparecida Gonzalez, Bruna Camila Quartim de Moraes, de Souza Rafael Queiroz, Graziano Kazuko Uchikawa, Lusignan Viktoriya. False positive results of Bowie and Dick type test used for hospital steam sterilizer with slower come-up ramps: A case study.. PloS one. 2020. DOI: 10.1371/journal.pone.0227943.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.