DOI 10.1016/j.joms.2019.11.013
In-House Surgeon-Led Virtual Surgical Planning for Maxillofacial Reconstruction.
Аннотация
PURPOSE: Virtual surgical planning (VSP) and custom fabricated cutting guides for maxillofacial reconstruction have been shown to improve the accuracy of bony reconstruction and overall surgical efficiency and decrease the ischemia time. Our aim was to describe an in-house VSP technique for maxillofacial reconstructive procedures.
MATERIALS AND
METHODS
We used 2 free software applications. 3DSlicer (available at: http://www.3dslicer.org) was used to extract the bones of interest for the recipient and the donor sites from the computed tomography scan's DICOM (digital imaging and communications in medicine) data. The Autodesk Meshmixer (Autodesk Inc, San Rafael, CA) was used to perform VSP and fabrication of the cutting guides. A reconstructed jaw model was printed in-house using a commercially available fused deposition modeling-based desktop 3-dimensional (3D) printer (Qidi Technology, Zhejiang, China) and used to prebend the reconstruction plate. The cutting guides were printed using a commercially available resin-based stereolithography apparatus desktop 3D printer (Form 2, Dental SG Resin; Formlabs, Somerville, MA) to allow for sterilization of the guides. We performed this technique for 19 consecutive patients with maxillofacial benign or malignant tumors requiring microvascular bony reconstruction. We calculated the average time and associated costs using this in-house VSP technique.
RESULTS
The technique was found to be simple and repeatable. The average time required for VSP was 158 minutes (2 hours, 38 minutes). The average cost for printing the reconstructed model per case was $5.21 Canadian dollars (CAD), and the average cost for printing the cutting guides per case was $12.80 CAD.
CONCLUSIONS
Using this technique, in-house VSP and 3D printing can be performed by the treating surgeon, without an engineering background, within a reasonable period.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.