[Quality Control of Medical Instruments and Cleaning Water].

Nakayama Hirobumi, Kawaguchi Kimiyuki, Kato Akihito, Morishima Hiroyuki, Nagashima Goro, Yagioka Setsuko, Asakura Takeshi, Tanaka Shinichi, Tanaka Yuichiro

СтатьяПеревод в очереди

Аннотация

PURPOSE: Herein we aimed to investigate the degradation of surgical instruments in our hospital and how water quality affects the rate of metal corrosion.

MATERIALS AND

METHODS

We observed 279 stainless steel instruments, and determined the presence of damage like metal corrosion or scale formation. We also measured the concentrations of chloride(Cl-)and silicate(SiO44-)ions in the water used for cleaning in our operating rooms, including tap water from the city water supply and reverse-osmosis(RO)filtered water.

RESULT: Pitting corrosion was observed on 71% of the instruments we investigated. The concentration of Cl- was 0.7mg/L in tap water and 0.1mg/L in RO water, while the concentration of SiO44- was 0.3mg/L in both the tap and RO water.

DISCUSSION: Of the dissolved ions Cl- and SiO44-, Cl- is more of a concern, as it causes pitting corrosion over time, while SiO44- causes scale formation. Considering the typical water quality in the operating-room environment, degradation must be monitored for the general maintenance of metal surgical instruments.

Для цитирования: Nakayama Hirobumi, Kawaguchi Kimiyuki, Kato Akihito, Morishima Hiroyuki, Nagashima Goro, Yagioka Setsuko, Asakura Takeshi, Tanaka Shinichi, Tanaka Yuichiro. [Quality Control of Medical Instruments and Cleaning Water].. No shinkei geka. Neurological surgery. 2019. DOI: 10.11477/mf.1436204035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.