MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00784-019-02926-9

Медицинская библиотека · Clinical oral investigations · 2019

Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing.

Pils Dieter, Baeppler Ralf J, Junker Rüdiger, Kielbassa Andrej M, Nothdurft Frank P

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Aim was to assess the influence of a standard autoclaving protocol on the retention of zirconia abutments glued on titanium bases in two-piece implant abutments.

MATERIALS AND

METHODS

Twenty-four CAD/CAM-generated zirconia abutments were adhesively cemented on prefabricated titanium bases. Before mechanical and thermal aging, the specimens were divided into two groups. Group 1 was subjected to a standard steam autoclaving protocol and Group 2 remained untreated (control). The tensile strength in all specimens was evaluated by a standardized pull-off test limited to a maximum force of 1000 N.

RESULTS

Eleven samples in both groups could be subjected to pull-off testing. Ten samples in Group 1 and three samples in Group 2 failed, while all others reached the maximum pull-off force. This difference was statistically significant. The mean retention values for the failed samples were 694.53 ± 369.10 N in Group 1 and 890.78 ± 25.90 N in Group 2. This difference was not statistically significant.

CONCLUSIONS

A standard autoclaving protocol does not reduce detachment force of two-piece zirconia abutments.

CLINICAL RELEVANCE: Clinical sterilization processes as recommended by regulatory authorities seem to be harmless to the structural integrity of two-piece zirconia implant abutments, at least with regard to the retention of the components.

Для цитирования: Pils Dieter, Baeppler Ralf J, Junker Rüdiger, Kielbassa Andrej M, Nothdurft Frank P. Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing.. Clinical oral investigations. 2019. DOI: 10.1007/s00784-019-02926-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.