Elevating the standard of endoscope processing: Terminal sterilization of duodenoscopes using a hydrogen peroxide-ozone sterilizer.

Molloy-Simard Vanessa, Lemyre Jean-Luc, Martel Karine, Catalone Bradley J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The health care community is increasingly aware of the processing challenges and infection risks associated with duodenoscopes owing to published reports of outbreaks and regulatory recalls. Studies have demonstrated that the current practices are inadequate for consistently producing patient-ready endoscopes. Alternatively, terminal sterilization would offer a greater margin of safety and potentially reduce the risk of patient infection. The purpose of this study was to evaluate the efficacy of a hydrogen peroxide-ozone sterilizer with regulatory clearance for terminal sterilization of duodenoscopes.

METHODS AND

RESULTS

Validation studies were performed under laboratory simulated-use and clinical in-use conditions. The overkill method study demonstrated a reduction of at least 6-log of Geobacillus stearothermophilus spores at half-cycle, providing a sterility assurance level of 10-6. In addition, the sterilizer achieved a 6-log reduction of G stearothermophilus in the presence of inorganic and organic soils in a simulated-use study. The clinical in-use study confirmed that the sterilizer achieved sterilization of patient-soiled duodenoscopes under actual use conditions.

CONCLUSIONS

Simulated-use and clinical in-use studies demonstrated the efficacy of a hydrogen peroxide-ozone sterilizer for terminal sterilization of duodenoscopes. This offers health care facilities a viable alternative for duodenoscope processing to enhance patient safety as part of a comprehensive infection control strategy.

Для цитирования: Molloy-Simard Vanessa, Lemyre Jean-Luc, Martel Karine, Catalone Bradley J. Elevating the standard of endoscope processing: Terminal sterilization of duodenoscopes using a hydrogen peroxide-ozone sterilizer.. American journal of infection control. 2018. DOI: 10.1016/j.ajic.2018.09.009.