DOI 10.1097/bot.0000000000001224
Effect of Repeated Sterilization by Different Methods on Strength of Carbon Fiber Rods Used in External Fixator Systems.
Аннотация
OBJECTIVES
We aimed to study the effects of repeated sterilization, using different methods, on the carbon fiber rods of external fixator systems.
METHODS
We used a randomized set of 44 unused, unsterilized, and identical carbon fiber rods (11 × 200 mm), randomly assigned to 2 groups: unsterilized (4 rods) and sterilized (40 rods). The sterilized rods were divided into 2 groups, those sterilized in an autoclave and those sterilized using hydrogen peroxide. These were further divided into 5 subgroups based on the number of sterilization repetition to which the fibers were subjected (25, 50, 75, 100, and 200). A bending test was conducted to measure the maximum bending force, maximum deflection, flexural strength, maximum bending moment and bending rigidity. We also measured the surface roughness of the rods.
RESULTS
An increase in the number of sterilization repetition led to a decrease in maximum bending force, maximum bending moment, flexural strength, and bending rigidity, but increased maximum deflection and surface roughness (P < 0.01). The effect of the number of sterilization repetition was more prominent in the hydrogen peroxide group.
CONCLUSIONS
This study revealed that the sterilization method and number of sterilization repetition influence the strength of the carbon fiber rods. Increasing the number of sterilization repetition degrades the strength and roughness of the rods.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.