Crawford Thomas C, Allmendinger Craig, Snell Jay, Weatherwax Kevin, Lavan Balasundaram, Baman Timir S, Sovitch Pat, Alyesh Daniel, Carrigan Thomas, Klugman Noah, Kune Denis, Hughey Andrew, Lautenbach Daniel, Sovitch Nathan, Tandon Karman, Samson George, Newman Charles, Davis Sheldon, Brown Archie, Wasserman Brad, Goldman Ed, Arlinghaus Sandra L, Oral Hakan, Eagle Kim A

Аннотация

OBJECTIVES: This study sought to develop a validated, reproducible sterilization protocol, which could be used in the reprocessing of cardiac implantable electronic devices (CIEDs).

BACKGROUND: Access to cardiac CIED therapy in high-income and in low- and middle-income countries varies greatly. CIED reuse may reduce this disparity.

METHODS: A cleaning and sterilization protocol was developed that includes washing CIEDs in an enzymatic detergent, screw cap and set screw replacement, brushing, inspection, and sterilization in ethylene oxide. Validation testing was performed to assure compliance with accepted standards.

RESULTS: With cleaning, the total mean bioburden for each of 3 batches of 10 randomly chosen devices was reduced from 754 to 10.1 colony-forming units. After sterilization with ethylene oxide, with 3 half-cycle and 3 full-cycle processes, none of the 90 biological indicator testers exhibited growth after 7 days. Through cleaning and sterilization, protein and hemoglobin concentrations were reduced from 99.2 to 1.42 μg/cm2 and from 21.4 to 1.03 μg/cm2, respectively. Mean total organic carbon residual was 1.44 parts per million (range 0.36 to 2.9 parts per million). Endotoxin concentration was not detectable at the threshold of <0.03 endotoxin units/ml or <3.0 endotoxin units/device. Cytotoxicity and intracutaneous reactivity tests met the standards set by the Association for Advancement of Medical Instrumentation and the International Organization for Standardization.

CONCLUSIONS: CIEDs can be cleaned and sterilized according to a standardized protocol achieving a 12-log reduction of inoculated product, resulting in sterility assurance level of 10-6.