DOI 10.1016/j.jmbbm.2018.01.033

Медицинская библиотека · Journal of the mechanical behavior of biomedical materials · 2018

Effects of sterilization on the mechanical properties of poly(methyl methacrylate) based personalized medical devices.

Münker T J A G, van de Vijfeijken S E C M, Mulder C S, Vespasiano V, Becking A G, Kleverlaan C J, CranioSafe Group, CranioSafe Group, Becking A G, Dubois L, Karssemakers L H E, Milstein D M J, van de Vijfeijken S E C M, Depauw P R A M, Hoefnagels F W A, Vandertop W P, Kleverlaan C J, Münker T J A G, Maal T J J, Nout E, Riool M, Zaat S A J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Nowadays, personalized medical devices are frequently used for patients. Due to the manufacturing procedure sterilization is required. How different sterilization methods affect the mechanical behavior of these devices is largely unknown.

MATERIALS AND

METHODS

Three poly(methyl methacrylate) (PMMA) based materials (Vertex Self-Curing, Palacos R+G, and NextDent C&B MFH) were sterilized with different sterilization methods: ethylene oxide, hydrogen peroxide gas plasma, autoclavation, and γ-irradiation. Mechanical properties were determined by testing the flexural strength, flexural modulus, fracture toughness, and impact strength.

RESULTS

The flexural strength of all materials was significantly higher after γ-irradiation compared to the control and other sterilization methods, as tested in a wet environment. NextDent C&B MFH showed the highest flexural and impact strength, Palacos R+G showed the highest maximum stress intensity factor and total fracture work.

CONCLUSION

Autoclave sterilization is not suitable for the sterilization of PMMA-based materials. Ethylene oxide, hydrogen peroxide gas plasma, and γ-irradiation appear to be suitable techniques to sterilize PMMA-based personalized medical devices.

Для цитирования: Münker T J A G, van de Vijfeijken S E C M, Mulder C S, Vespasiano V, Becking A G, Kleverlaan C J, CranioSafe Group, CranioSafe Group, Becking A G, Dubois L, Karssemakers L H E, Milstein D M J, van de Vijfeijken S E C M, Depauw P R A M, Hoefnagels F W A, Vandertop W P, Kleverlaan C J, Münker T J A G, Maal T J J, Nout E, Riool M, Zaat S A J. Effects of sterilization on the mechanical properties of poly(methyl methacrylate) based personalized medical devices.. Journal of the mechanical behavior of biomedical materials. 2018. DOI: 10.1016/j.jmbbm.2018.01.033.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.