DOI 10.1016/j.ajic.2017.07.021
Improper positioning of the elevator lever of duodenoscopes may lead to sequestered bacteria that survive disinfection by automated endoscope reprocessors.
Аннотация
BACKGROUND
Some outbreaks associated with contaminated duodenoscopes have been attributed to biofilm formation. The objective of this study was to determine whether bacteria within an organic matrix could survive if the elevator lever was improperly positioned in the automated endoscope reprocessor (AER) after 1 round of reprocessing.
METHODS
Duodenoscope lever cavities with an open or sealed elevator wire channel were inoculated with 6-7 Log10 of both Escherichia coli and Enterococcus faecalis in ATS2015 (Healthmark Industries, Fraser, MI) and dried for 2 hours. The duodenoscopes with the lever in the horizontal position were processed through 2 makes of AERs. The cavity was sampled using a flush-brush-flush method to determine the quantity of surviving bacteria.
RESULTS
E faecalis (range, 21-6 Log10 CFU) and E coli (range, 0-3 Log10 CFU) survived disinfection of sealed or unsealed elevator wire channel duodenoscopes in 2 different AERs with and without cleaning cycles.
CONCLUSION
If bacteria in organic residue are under the improperly positioned lever, then just 1 round of use is sufficient for bacteria to survive both liquid chemical sterilization and liquid chemical HLD regardless of whether or not the AER had a cleaning cycle.