MedElevator Войти в исследования
← Поиск публикаций

DOI 10.28920/dhm53.4.327-332

Медицинская библиотека · Diving and hyperbaric medicine · 2023

Comparing the EMMA capnograph with sidestream capnography and arterial carbon dioxide pressure at 284 kPa.

Vrijdag Xavier Ce, van Waart Hanna, Sames Chris, Sleigh Jamie W, Mitchell Simon J

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Capnography aids assessment of the adequacy of mechanical patient ventilation. Physical and physiological changes in hyperbaric environments create ventilation challenges which make end-tidal carbon dioxide (ETCO2) measurement particularly important. However, obtaining accurate capnography in hyperbaric environments is widely considered difficult. This study investigated the EMMA capnograph for hyperbaric use.

METHODS

We compared the EMMA capnograph to sidestream capnography and the gold standard arterial carbon dioxide blood gas analysis in a hyperbaric chamber. In 12 resting subjects breathing air at 284 kPa, we recorded ETCO2 readings simultaneously derived from the EMMA and sidestream capnographs during two series of five breaths (total 24 measurements). An arterial blood gas sample was also taken simultaneously in five participants.

RESULTS

Across all measurements there was a difference of about 0.1 kPa between the EMMA and sidestream capnographs indicating a very slight over-estimation of ETCO2 by the EMMA capnograph, but fundamentally good agreement between the two end-tidal measurement methods. Compared to arterial blood gas pressure the non-significant difference was about 0.3 and 0.4 kPa for the EMMA and sidestream capnographs respectively.

CONCLUSIONS

In this study, the EMMA capnograph performed equally to the sidestream capnograph when compared directly, and both capnography measures gave clinically acceptable estimates of arterial PCO2.

Для цитирования: Vrijdag Xavier Ce, van Waart Hanna, Sames Chris, Sleigh Jamie W, Mitchell Simon J. Comparing the EMMA capnograph with sidestream capnography and arterial carbon dioxide pressure at 284 kPa.. Diving and hyperbaric medicine. 2023. DOI: 10.28920/dhm53.4.327-332.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.