Intraoperative Factors that Predict the Successful Placement of Essure Microinserts.

Arthuis Chloé J, Simon Emmanuel G, Hébert Thomas, Marret Henri

СтатьяПеревод в очереди

Аннотация

STUDY

OBJECTIVE

To determine whether the number of coils visualized in the uterotubal junction at the end of hysteroscopic microinsert placement predicts successful tubal occlusion.

DESIGN

Cohort retrospective study (Canadian Task Force classification II-2).

SETTING

Department of obstetrics and gynecology in a teaching hospital.

PATIENTS: One hundred fifty-three women underwent tubal microinsert placement for permanent birth control from 2010 through 2014. The local institutional review board approved this study.

INTERVENTION: Three-dimensional transvaginal ultrasound (3D TVU) was routinely performed 3 months after hysteroscopic microinsert placement to check position in the fallopian tube.

MEASUREMENTS AND MAIN

RESULTS

The correlation between the number of coils visible at the uterotubal junction at the end of the hysteroscopic microinsert placement procedure and the device position on the 3-month follow-up 3D TVU in 141 patients was evaluated. The analysis included 276 microinserts placed during hysteroscopy. The median number of coils visible after the hysteroscopic procedure was 4 (interquartile range, 3-5). Devices for 30 patients (21.3%) were incorrectly positioned according to the 3-month follow-up 3D TVU, and hysterosalpingography was recommended. In those patients the median number of coils was in both the right (interquartile range, 2-4) and left (interquartile range, 1-3) uterotubal junctions. The number of coils visible at the uterotubal junction at the end of the placement procedure was the only factor that predicted whether the microinsert was well positioned at the 3-month 3D TVU confirmation (odds ratio, .44; 95% confidence interval, .28-.63). When 5 or more coils were visible, no incorrectly placed microinsert could be seen on the follow-up 3D TVU; the negative predictive value was 100%. No pregnancies were reported.

CONCLUSION

The number of coils observed at the uterotubal junction at the time of microinsert placement should be considered a significant predictive factor of accurate and successful microinsert placement.

Для цитирования: Arthuis Chloé J, Simon Emmanuel G, Hébert Thomas, Marret Henri. Intraoperative Factors that Predict the Successful Placement of Essure Microinserts.. Journal of minimally invasive gynecology. 2017. DOI: 10.1016/j.jmig.2017.02.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.