American journal of infection control · 2017
An economic analysis of the benefits of sterilizing medical instruments in low-temperature systems instead of steam.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
McCreanor Victoria, Graves Nicholas
Аннотация
BACKGROUND: Hydrogen peroxide-based, low-temperature sterilization has been shown to do less damage to medical instruments than steam autoclaves. However, low-temperature systems are more expensive to run. Higher costs need to be balanced against savings from reduced repair costs to determine value for money when choosing how to sterilize certain instruments, which are able to be reprocessed in either system.
METHODS: This analysis examines the economic effects of using low-temperature sterilization systems to reprocess rigid and semi-rigid endoscopes, which are sensitive to heat and moisture, but still able to be sterilized using steam. It examines the changes to costs and frequency of repairs expected over 10 years, resulting from a choice to sterilize these instruments in a low-temperature system instead of steam.
RESULTS: Overall, the results showed that increased sterilization costs are outweighed by the savings associated with less frequent repairs. Over a 10-year period, in large health care facilities, the probability of achieving an internal rate of return of at least 6% is 0.81.
CONCLUSIONS: Our model shows it is likely to be a good decision for large health care facilities to invest in low-temperature sterilization systems.
METHODS: This analysis examines the economic effects of using low-temperature sterilization systems to reprocess rigid and semi-rigid endoscopes, which are sensitive to heat and moisture, but still able to be sterilized using steam. It examines the changes to costs and frequency of repairs expected over 10 years, resulting from a choice to sterilize these instruments in a low-temperature system instead of steam.
RESULTS: Overall, the results showed that increased sterilization costs are outweighed by the savings associated with less frequent repairs. Over a 10-year period, in large health care facilities, the probability of achieving an internal rate of return of at least 6% is 0.81.
CONCLUSIONS: Our model shows it is likely to be a good decision for large health care facilities to invest in low-temperature sterilization systems.